Product Overview
The HORIBA Yumizen C560 Clinical Chemistry Analyzer is a fully automated, floor-standing system designed for midsize clinical laboratories. It combines random-access chemistry testing, STAT sample priority and automated sample and reagent management in a single platform.
The analyzer processes up to 400 photometric tests per hour and supports up to 240 ISE tests per hour when configured with the ISE module. A 90-position sample tray, 80 refrigerated reagent positions and permanent reaction cuvettes support routine chemistry workflows while helping laboratories manage sample and reagent volume.
Jant catalog number HORYUMIZENC560 identifies the current listing. Contact Jant Pharmacal to confirm current availability, the included ISE configuration, reagent menu, computer and software, LIS requirements, installation, training, warranty and service support.
400 Photometric Tests per Hour
Automated random-access operation with STAT sample priority for routine chemistry testing.
90 Sample Positions
Accommodates primary tubes, secondary tubes and sample cups with barcode identification.
80 Refrigerated Reagent Positions
Onboard refrigeration at 2-10°C supports reagent storage during analyzer operation.
Key Features
- Fully automated, discrete, random-access clinical chemistry analyzer
- STAT sample priority capability
- Up to 400 photometric tests per hour
- Up to 240 ISE tests per hour when configured for sodium, potassium and chloride testing
- Absorbance photometry and turbidimetry measurement principles
- End-point, fixed-time and kinetic methodologies
- Support for single-, dual-, triple- and quadruple-reagent chemistries
- 90 sample positions for primary tubes, secondary tubes and sample cups
- 80 reagent positions in a refrigerated compartment
- 90 permanent reaction cuvettes with automated washing
- Liquid-level, clot, bubble and collision detection
- Automated pre-dilution and post-dilution up to 1:150
- Sample and reagent barcode reading
- Real-time analytical, carousel and quality-control status monitoring
- Bi-directional LIS interface capability
- Front-access maintenance area with guided maintenance functions
Designed for Efficient Chemistry Workflows
Sample and Probe Management
Barcode identification, automated dilution, STAT priority and probe safeguards support efficient handling of routine and priority samples.
Reagent and Reaction Management
Refrigerated reagent storage, inventory checks, low-volume reactions and an eight-stage cuvette wash station support routine operation.
QC and Connectivity
Levey-Jennings monitoring, Westgard multi-rule support, reagent-lot QC and bi-directional LIS capability assist with results review and traceability.
Technical Specifications
| Specification | HORIBA Yumizen C560 |
|---|---|
| Jant Catalog Number | HORYUMIZENC560 |
| System Type | Fully automated, discrete, random-access floor-standing analyzer with STAT priority |
| Measurement Principles | Absorbance photometry and turbidimetry; ISE when configured |
| Photometric Throughput | Up to 400 tests per hour |
| ISE Throughput | Up to 240 tests per hour; confirm the included ISE configuration |
| ISE Parameters | Sodium, potassium and chloride |
| Sample Capacity | 90 positions for primary tubes, secondary tubes and sample cups |
| Sample Volume | 1.5-50 µL in 0.1 µL increments |
| Automatic Dilution | Pre-dilution and post-dilution up to 1:150 |
| Reagent Capacity | 80 positions in a refrigerated 2-10°C compartment |
| Reaction System | 90 permanent cuvettes with automated washing; two independent mixers |
| Reaction Volume | 120-360 µL |
| Wavelengths | 340, 380, 412, 450, 505, 546, 570, 605, 660, 700, 740 and 800 nm |
| Barcode Reading | Sample and reagent barcode reading |
| Connectivity | Bi-directional LIS capability; RS-232, network and USB/parallel interfaces |
| Water Requirement | Approximately 20 L/hour of deionized water; confirm the installation configuration |
| Dimensions | Approximately 46 W × 28 D × 45 H inches |
| Weight | Approximately 661 lb |
Specifications and available configurations may change. Confirm the analyzer and ISE configuration, computer and software, reagent menu, water and waste connections, electrical service, LIS components, installation requirements and service coverage with Jant before ordering.
Complete the Clinical Chemistry Workflow
Connect the Yumizen C560 with the appropriate chemistry reagents, calibrators and quality-control materials for the laboratory’s validated test menu.
Quality Controls
Explore quality-control materials for chemistry and related laboratory workflows.
Clinical Chemistry
Browse Jant’s clinical chemistry analyzers, reagents and related laboratory products.
Documentation and Resources
HORIBA Yumizen C560 Product BrochureUse current manufacturer instructions, assay-specific labeling and validated application parameters. Contact Jant for current manuals, installation requirements and reagent documentation.
Frequently Asked Questions
The Yumizen C560 processes up to 400 photometric tests per hour. HORIBA also lists ISE throughput of up to 240 tests per hour when the analyzer is configured for sodium, potassium and chloride testing.
The Yumizen C560 supports ISE testing for sodium, potassium and chloride, but the current Jant listing does not clearly state whether the module is included with catalog HORYUMIZENC560. Confirm the quoted configuration before ordering.
The sample tray provides 90 positions for primary tubes, secondary tubes and sample cups. The system supports barcode identification and STAT sample priority.
The analyzer provides 80 positions in a refrigerated compartment maintained at 2-10°C. Confirm the current reagent menu, bottle configuration, calibrators and controls with Jant.
The current U.S. brochure lists sample volumes of 1.5-50 µL and reaction volumes of 120-360 µL. Required volumes vary by application and should be confirmed using the current assay instructions.
The platform supports bi-directional LIS communication. Confirm the required interface, computer, middleware, mapping and validation process for your laboratory environment.
Confirm the analyzer and ISE configuration, electrical service, deionized-water and waste requirements, floor space and service clearances, computer and software, LIS connection, delivery access, installation, training and service coverage.
Ordering Information
To confirm availability or request a quote, call Jant Pharmacal at 800.676.5565
| Catalog Number | Description | Unit of Measurement |
|---|---|---|
| HORYUMIZENC560 | HORIBA Yumizen C560 Clinical Chemistry Analyzer | Each |
Confirm the ISE configuration, reagent menu, calibrators, controls, computer and LIS components, installation, training, warranty and service coverage when requesting a quote.